Technical data
2067634-131 Revision E
SEER ™ 1000
119
General specifications table 202
Policies and MAKER statement - RESISTANCE TO INTERFERENCE ELECTROMAGNETIC
The device is designed for use in the electromagnetic environment specified below. The customer or the device user
must ensure that it is used in such an environment.
test
electromagnetic
interference
Test level IEC
60601
matching level
ELECTROMAGNETIC ENVIRONMENT
DIRECTIVES
discharge
Electrostatic (DE)
according to IEC
61000-4-2
•
Contact
discharge 6 kV
•
8 kV air discharge
•
Contact
discharge 6 kV
•
8 kV air discharge
Floors should be wood, concrete or ceramic. If
floors are covered with synthetic material, the
relative humidity should be at least 30%.
rapid and transient
quantities of
electrical
interference / IEC
61000-4-4 Bursts
as
•
2 kV network
cables
•
1 kV for cable entry
and exit
Not applicable
The supply voltage should correspond to the
quality of a typical commercial or hospital
environment.
Transient voltage
(Surges) according to
IEC 61000-4-5
•
Voltage of 1 kV between
the external-conductive
external conductor
•
Voltage of 2 kV between
the external ground
conductor,
Not applicable
The supply voltage should correspond to the
quality of a typical commercial environment or a
hospital environment.
Voltage
drops,
short interruptions and
variations in the supply
voltage according to
IEC 61000-4-11
<5% UT for ½ period
(> 95% decrease)
40% UT for 5 times
(60% decrease) 70%
UT for 25 peroxides
(30% decrease) <5%
UT for 5 s (> 95%
which gives)
Not applicable
The quality of the supply voltage should
correspond to a typical feed or a commercial
environment of a hospital environment. If the
device user require a continuous FUNCTION,
even during power supply interruptions, it is
recommended that the device be powered from
an uninterruptible power supply or a battery.