Problems solution
2067634-131 Revision E
SEER ™ 1000
103
Message
Recommendation
"Battery level too low, recorder is not turned off."
Insert a new disposable alkaline battery or an
NiMH battery that is sufficiently charged.
"The recorder is recording, you want to select a
recorder differently?"
a recording is already in progress. Stop and
delete the recording, or select a different
recorder.
"The recorder is not connected to this device. Please
repeat the coupling to this recorder. "
(Apple iOS Post)
1.
Close the application.
2.
Call the "Settings" menu.
3.
Select the "Bluetooth" tab.
4.
Select the "Ignore this device" option.
5.
Select the recorder.
6.
Repeat coupling to the recorder.
"The data are stored on the recorder, you wish to
select a different recorder?"
Erasing the data stored on the recorder or
transfer them to a long-term ECG analysis
software. You can also select a different
recorder.
"Incorrect password."
The password is incorrect. Try a valid password.
"Technical wrong or bad password."
Technical ID does not exist or the password is
incorrect. Try a combination of valid technician ID
and password.