Technical data
120
SEER ™ 1000
2067634-131 Revision E
Policies and MAKER statement - RESISTANCE TO INTERFERENCE ELECTROMAGNETIC
Field
Magnetic with power
frequency (50/60 Hz)
IEC 61000-4-8
3 A / m
Not applicable
If there are faults it may be necessary to place
the device further away from sources of
magnetic fields of power frequency or install
magnetic shielding: the magnetic field energy
frequency must be measured at the designated
installation location, to ensure that it is
sufficiently small.
NOTE: UT is the mains alternating voltage before application of the test level.