Problems solution
98
SEER ™ 1000
2067634-131 Revision E
error signal
Cause
Recommendation
At
LED event keys
At
LED event keys
They are lit in yellow.
A previous recording or an entry
from a previous patient was not
cleared.
Delete a previous recording record or
previous recordings or previous data
of the patient. Use to this, during
ECG analysis software prazoou
SEER 1000 App.
O
LED event button
O
LED event button
will stop flashing green during
the set recording period.
The recording was
automatically stopped before
reaching the duration of the
scheduled recording.
Possibly the recording was started
with a low battery. Save or delete the
recording and insert a new battery to
start a new recording. Make sure the
machine is damaged. Please contact
the customer service center if the
device is damaged.
At
LED event keys
At
LED event keys
and
LED electrodes
and
LED electrodes
flash three times per
second,
alternately in green and
yellow.
Recording can not be started. The
self-test was not successful.
More details can be found in the "
How
More details can be found in the "
How