![FujiFilm SONOSITE X-PORTE Скачать руководство пользователя страница 216](http://html1.mh-extra.com/html/fujifilm/sonosite-x-porte/sonosite-x-porte_user-manual_2342271216.webp)
7-6
Troubleshooting and maintenance
Patient data
Digital Imaging and Communications in Medicine (DICOM) provides a way to archive patient data by
connecting your ultrasound system over a local area network (LAN) with various archivers for storage after
every patient exam. FUJIFILM SonoSite recommends that you configure and use DICOM transfer to prevent
patient data loss in the event of a system fault. For more information, see
If you do not use DICOM networking, then FUJIFILM SonoSite recommends that you configure your system
for automatic export of patient data to a USB memory stick after every exam. For more information, see
System configuration settings
In addition to patient data, FUJIFILM SonoSite recommends that you back up ultrasound system
configuration settings, called presets, after you have fully configured the system and any time you modify
these settings. These backups preserve your customized settings in case of a fault in the system. For more
information, see
Servicing
Your ultrasound engine may be repaired or replaced at the manufacturer’s discretion. If servicing is
necessary, then you must remove the ultrasound engine. Before the engine is removed and shipped to a
repair facility, you must take precautions to protect patient data and to preserve your customized settings.
To prepare your system for service
1
End any in-progress procedures.
2
Export all patient procedure information to a USB stick or archive it to a DICOM device. For complete
instructions, see
3
To delete all patient data, tap
Patient
and then
Patient List
.
4
Tap
Select All
and then tap
Delete
.
5
If using DICOM, delete worklist data by tapping
Patient
and then
Worklist
. Tap
Clear
.
6
Export the following items to a USB memory stick:
System preferences (that is, presets)
Cautions
To protect patient privacy, all patient procedure information must be exported to
a USB memory stick or archived to a secure repository via DICOM transfer and
then deleted from the Patient List.
To preserve your configuration settings, export Presets and DICOM settings to a
USB memory stick and store the stick in a secure location.
Содержание SONOSITE X-PORTE
Страница 1: ...USER GUIDE ...
Страница 75: ...Setting up the system 3 39 ...
Страница 76: ...3 40 Setting up the system ...
Страница 128: ...4 52 Imaging ...
Страница 186: ...5 58 Measurements and calculations ...
Страница 210: ...6 24 Measurement references ...
Страница 234: ...8 16 Cleaning and disinfecting ...
Страница 262: ...9 28 Safety ...
Страница 334: ...10 72 Acoustic output ...
Страница 364: ...B 14 Index printing 5 56 zoom 4 23 ...
Страница 365: ......
Страница 366: ...P14645 05 P14645 05 ...