8-8
Cleaning and disinfecting
Clean and disinfect system and transducer to a low level
(non-critical uses)
Use the following procedure to clean and disinfect the ultrasound system and transducer
if it has not come
into contact with blood, broken skin, mucosal membranes, or bodily fluids
.
Follow the manufacturer's instructions when using cleaners and disinfectants. The cleaners and disinfectants
listed in the procedure are both chemically compatible and have been tested for efficacy with the system
and transducers. Confirm that the cleaners and disinfectants are appropriate for your facility’s use.
To clean and disinfect the system and transducer
1 Turn off
the system by pressing the
Power button.
2 Unplug
the power cord from the outlet.
3 Remove
the transducer sheath, if applicable.
WARNING
If the system or transducer has come into contact with any of the following, use
the high-level cleaning and disinfection procedure. See
system and transducer to a high level (semi-critical uses)”
Blood
Broken skin
Mucosal membranes
Bodily fluids
WARNINGS
To avoid electrical shock, before cleaning, turn off the system and disconnect it
from the power supply.
Wear the appropriate personal protective equipment (PPE) recommended by
the chemical manufacturer, such as eyewear and gloves.
Cautions
Do not spray cleaners or disinfectants directly on the system surfaces or on
system and transducer connectors. Doing so may cause solution to leak into the
system, damaging it and voiding the warranty.
Use only FUJIFILM SonoSite-recommended cleaners and disinfectants. Using a
non- recommended disinfecting solution or incorrect solution strength can
damage the System and transducer and void the warranty. Follow the
disinfectant manufacturer’s recommendations for solutions strengths.
Do not attempt to disinfect a transducer or transducer cable using a method or
chemical that is not included here. Doing so can damager the transducer and
void the warranty.
B
Option
Содержание SONOSITE X-PORTE
Страница 1: ...USER GUIDE ...
Страница 75: ...Setting up the system 3 39 ...
Страница 76: ...3 40 Setting up the system ...
Страница 128: ...4 52 Imaging ...
Страница 186: ...5 58 Measurements and calculations ...
Страница 210: ...6 24 Measurement references ...
Страница 234: ...8 16 Cleaning and disinfecting ...
Страница 262: ...9 28 Safety ...
Страница 334: ...10 72 Acoustic output ...
Страница 364: ...B 14 Index printing 5 56 zoom 4 23 ...
Страница 365: ......
Страница 366: ...P14645 05 P14645 05 ...