IT Network
11-3
IT network failure recovery measures
Connection to an IT network may become, at times, unreliable, and this may lead to failure to perform the
functions described in
on page 11-1. As a result, the following hazardous situations may occur:
Table 11-1: IT network failure recovery
Network failure
Impact on equipment
Hazard
SonoSite X-Porte
countermeasures
IT network becomes
unstable
Unable to transmit
exam data to a PACS
Delay of diagnosis
SonoSite X-Porte has
internal memory, and
exam data is stored in
it. After the IT network
has returned to stable,
the user can re-initiate
the transfer of data.
Delay of transmission
to a PACS
Incorrect data
transmitted to a PACS
Misdiagnosis
Integrity of the data is
ensured by the TCP/IP
and DICOM Protocols
used by SonoSite
X-Porte.
Unable to get order
data from an MWL
server
Delay of exam
On SonoSite X-Porte,
the user can initiate/
create a new study.
Delay of getting order
from an MWL server
Incorrect data from a
MWL server
Incorrect exam
SonoSite X-Porte uses
the TCP/IP and DICOM
Protocols. Integrity of
the data is ensured by
them.
Unable to get the time
from a time server.
Incorrect exam data
SonoSite X-Porte has
the capability of
entering data and time
manually.
Incorrect time data
SonoSite X-Porte
always indicates the
date and the time on
the main screen.
Содержание SONOSITE X-PORTE
Страница 1: ...USER GUIDE ...
Страница 75: ...Setting up the system 3 39 ...
Страница 76: ...3 40 Setting up the system ...
Страница 128: ...4 52 Imaging ...
Страница 186: ...5 58 Measurements and calculations ...
Страница 210: ...6 24 Measurement references ...
Страница 234: ...8 16 Cleaning and disinfecting ...
Страница 262: ...9 28 Safety ...
Страница 334: ...10 72 Acoustic output ...
Страница 364: ...B 14 Index printing 5 56 zoom 4 23 ...
Страница 365: ......
Страница 366: ...P14645 05 P14645 05 ...