Acoustic output
10-59
Table 10-47: Transducer Model: P10xp, Operating Mode: CW Doppler
Index Label
M.I.
TIS
TIB
TIC
Scan
Non-scan
Non-scan
A
aprt
≤1
A
aprt
>1
Global Maximum Index Value
(a)
-
(a)
-
2.4
1.5
Associated
Acoustic
Parameter
p
r.3
(MPa)
#
W
0
(mW)
-
#
34.3
37.4
min of [W
.3
(z
1
),I
TA.3
(z
1
)] (mW)
-
z
1
(cm)
-
Z
bp
(cm)
-
Z
sp
(cm)
0.50
z@Pll
.3max
(cm)
#
d
eq
(Z
sp
)
(cm)
0.28
F
c
(MHz)
#
-
#
-
4.00
4.00
Dim of A
aprt
X (cm)
-
#
-
0.42
0.45
Y (cm)
-
#
-
0.70
0.70
Other
Information
PD
(μsec)
#
PRF
(Hz)
#
p
r
@PII
max
(MPa)
#
d
eq
@Pll
max
(cm)
0.28
Focal Length
FL
x
(cm)
-
#
-
1.54
FL
y
(cm)
-
#
-
5.00
I
PA.3
@MI
max
(W/cm
2
) #
Operating
Control
Conditions
Control 1: Exam Type
Crd
Crd
Control 2: Sample Volume
Position
Zone 1
(9 mm)
Zone 2
(15 mm)
(a) This index is not required for this operating mode; value is <1.
(b) This transducer is not intended for transcranial or neonatal cephalic uses.
# No data are reported for this operating condition since the global maximum index value is not reported for the
reason listed. (Reference Global Maximum Index Value line.)
- Data are not applicable for this transducer/mode.
Содержание SONOSITE X-PORTE
Страница 1: ...USER GUIDE ...
Страница 75: ...Setting up the system 3 39 ...
Страница 76: ...3 40 Setting up the system ...
Страница 128: ...4 52 Imaging ...
Страница 186: ...5 58 Measurements and calculations ...
Страница 210: ...6 24 Measurement references ...
Страница 234: ...8 16 Cleaning and disinfecting ...
Страница 262: ...9 28 Safety ...
Страница 334: ...10 72 Acoustic output ...
Страница 364: ...B 14 Index printing 5 56 zoom 4 23 ...
Страница 365: ......
Страница 366: ...P14645 05 P14645 05 ...