![FujiFilm SONOSITE X-PORTE Скачать руководство пользователя страница 192](http://html1.mh-extra.com/html/fujifilm/sonosite-x-porte/sonosite-x-porte_user-manual_2342271192.webp)
6-6
Measurement references
E/Ea Ratio
Reynolds, Terry.
The Echocardiographer’s Pocket Reference
. 2nd ed., School of Cardiac Ultrasound,
Arizona Heart Institute, (2000), p.225.
E Velocity/Ea velocity
where:
E velocity = Mitral Valve E velocity
Ea = annular E velocity, also known as E prime
Effective Regurgitant Orifice (ERO) in mm
2
Oh, J.K., J.B. Seward, A.J. Tajik.
The Echo Manual
. 3rd ed., Lippincott, Williams, and Wilkins, (2007),
p.73-76, p.210.
ERO = MV Flow Rate/ MR Vel * 100
Ejection Fraction (EF), percent
Oh, J.K., J.B. Seward, A.J. Tajik.
The Echo Manual
.3rd ed., Lippincott, Williams, and Wilkins, (2007),
p.115-116.
EF = [(LVEDV – LVESV)/LVEDV)] * 100%
where:
EF = Ejection Fraction
LVEDV = Left Ventricular End Diastolic Volume
LVESV = Left Ventricular End Systolic Volume
Elapsed Time (ET) in msec
ET = time between velocity cursors in milliseconds
FAC (LV)
Dennis, A.T. et al. “Haemodynamics in women with untreated pre-eclampsia.”
Anaesthesia
. (2012), 67:
p.1105-1118.
LV FAC (%) = (LV EDA – LV ESA) / LV EDA * 100
where:
LV FAC = Left Ventricular Fractional Area Change (%)
LV EDA = Left Ventricular End-diastolic area (cm^2)
LV ESA = Left Ventricular End-systolic area (cm^2)
Содержание SONOSITE X-PORTE
Страница 1: ...USER GUIDE ...
Страница 75: ...Setting up the system 3 39 ...
Страница 76: ...3 40 Setting up the system ...
Страница 128: ...4 52 Imaging ...
Страница 186: ...5 58 Measurements and calculations ...
Страница 210: ...6 24 Measurement references ...
Страница 234: ...8 16 Cleaning and disinfecting ...
Страница 262: ...9 28 Safety ...
Страница 334: ...10 72 Acoustic output ...
Страница 364: ...B 14 Index printing 5 56 zoom 4 23 ...
Страница 365: ......
Страница 366: ...P14645 05 P14645 05 ...