Cleaning and disinfecting
8-15
Cleaning and disinfecting the ECG cable and slave cable
To clean and disinfect the ECG cables (wipe method)
1
Remove the cable from the system.
2
Examine the ECG cable for damage such as cracks or splitting.
3
Clean the surface using a soft cloth lightly dampened in a mild soap, cleaning solution, or pre-moistened
wipe. Apply the solution to the cloth rather than the surface.
4
Wipe the surfaces with a FUJIFILM SonoSite Approved cleaner or disinfectant.
5
Air dry or dry with a clean cloth.
For more information on the ECG slave cable, refer to the
ECG Slave Cable User Guide
.
Caution
To avoid damaging the ECG cable, do not sterilize.