9-12
Safety
Prevent the buildup of electrostatic charge. For example, use humidification, conductive floor coverings,
nonsynthetic clothing, ionizers, and minimizing insulating materials.
Discharge your body to earth.
Use a wrist strap to bond yourself to the ultrasound system or to earth.
Separation distance
Recommended separation distances between portable and mobile RF communications
equipment and the SonoSite X-Porte ultrasound system
The SonoSite X-Porte ultrasound system is intended for use in an electromagnetic environment in which
radiated radio frequency (RF) disturbances are controlled. The customer or the user of the SonoSite X-Porte
ultrasound system can help prevent electromagnetic interference by maintaining a minimum distance
between portable and mobile RF communications equipment (transmitters) and the SonoSite X-Porte
ultrasound system as recommended below, according to the maximum output power of the
communications equipment.
Compatible accessories and peripherals
FUJIFILM SonoSite has tested the SonoSite X-Porte ultrasound system with the following accessories and
peripherals and has demonstrated compliance to the requirements of IEC 60601-1-2:2007.
Table 9-1: Separation distance
Rated maximum
output power
of transmitter
(Watts)
Separation distance according to frequency of transmitter (m)
150 kHz to 80 MHz
d=1.2 √P
80 MHz to 800 MHz
d=1.2 √P
800 MHz to 2.5 GHz
d=2.3 √P
0.01
0.12
0.12
0.23
0.1
0.38
0.38
0.73
1
1.2
1.2
2.3
10
3.8
3.8
7.3
100
12
12
23
For transmitters rated at a maximum output power not listed above, the recommended separation distance (d) in
meters (m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the
maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.
Содержание SONOSITE X-PORTE
Страница 1: ...USER GUIDE ...
Страница 75: ...Setting up the system 3 39 ...
Страница 76: ...3 40 Setting up the system ...
Страница 128: ...4 52 Imaging ...
Страница 186: ...5 58 Measurements and calculations ...
Страница 210: ...6 24 Measurement references ...
Страница 234: ...8 16 Cleaning and disinfecting ...
Страница 262: ...9 28 Safety ...
Страница 334: ...10 72 Acoustic output ...
Страница 364: ...B 14 Index printing 5 56 zoom 4 23 ...
Страница 365: ......
Страница 366: ...P14645 05 P14645 05 ...