36
High
Upstream occlusion ,
detection when infusing
only
Flow obstruction above the
pump
Tube is bent between
container and pump
If applicable, roller clamp
closed
Blockage in bag or spike
yes
alarm
Remove kink
Open roller clamp
Check bag and spike
Note: the upstream occlusion
alarm becomes active only after 2
ml of infusion solution is pumped
successfully. A set that is occluded
when loaded in the pump will not
trigger this alarm.*
High
Line empty
Container empty
after 0.75 ml of continuous
air infusion
Set not fully primed
Air sensor area
contaminated
Set inserted incorrectly
yes
alarm
Stop infusion or change container
and set
Prime set
Clean the air sensor area using a
lint-free cloth lightly dampened
with warm water and a mild
detergent or disinfectant or follow
the recommended procedure.
Allow the pump to dry. For
cleaning procedures see Cleaning
Check the position of the set and
insert correctly.
High
Batteries are defective
(not charging correctly)
Pump is plugged in at the
wall outlet
no
alarm
Contact service *
Low
Battery pre-alarm
no
warning
Battery will be depleted in
30 minutes. Plug pump into the
wall outlet
High
Battery depleted
No electrical power
yes
alarm
No external power available.
Check proper installation in the
holder.
Check contact pins of holder and
pump.
Check cable and connection to the
wall outlet.
Battery is nearly depleted. Plug
into the wall outlet to continue
infusion or stop infusion.*
High
Door open
Pump door or lever open
Pump door installed
incorrectly
Door mechanism defective
yes
alarm
Close pump door and lever
Remove door and reinstall
Contact service*
Blinking
Symbols
Alarm
Priority
Meanings
Pumping
stop
Type
Operating
conditions/Actions