9
In case of refrigerated products, allow the product to reach the
operating temperature range before use.
When the device is stored at the minimum temperature of -20°C,
wait for 2 hours to allow the product to reach the operating
temperature range before using the pump. False occlusion alarm
can be triggered if the pump/set temperature is too low.
When the device is stored at the maximum temperature of
+45°C, it is ready to be used at ambient temperature.
Warning
This device can be affected by environmental pressure or
pressure variations, mechanical shocks, heat ignition sources,
radio-frequency interference (RFI) or electromagnetic radiation.
The pump should only be connected to the wall outlet with the
power cord supplied by the manufacturer.
Check that the line power voltage corresponds with the value
indicated on the device label.
Do not exceed the permitted voltage on the external connections.
A non-medical electric device connected to the device must
conform to the suitable IEC/EN standard (e.g. IEC/EN 60950 and
IEC/EN 60601-1).
Fresenius Kabi
accepts no liability for use of any interface
communication between the
Ambix
activ
and computer
systems.
The pump should be used under specified operational, storage
and transport conditions, see Section 7.2, page 43.
Use of ASV (Anti Siphoning Valve):
-
Ambix
activ
may be used with any ASV with a triggering
pressure lower or equal than 0.2 bar.
- ASV will affect the flow rate with a negative bias
≤
3%.