![Fresenius Kabi Ambix activ Скачать руководство пользователя страница 24](http://html.mh-extra.com/html/fresenius-kabi/ambix-activ/ambix-activ_instructions-for-use-manual_2331619024.webp)
24
3.5
Programming the Ambix
activ
3.5.1
Continuous Mode
3.5.2
Volume/Time Mode
3.5.3
Intermittent Mode
Button
Action
ON/OFF
Switch pump on
MODE
Change to continuous mode
+/–
Set rate
START
Start pumping
Information
Press the + and – buttons in the lowest row simultaneously to
activate the target volume.
To abort the program, switch pump off.
Button
Action
ON/OFF
Switch pump on
MODE
Change to volume/time mode
Middle row +/–
Set application time
Lower row +/–
Set target volume
START
Start pumping
Information
The flow rate will be calculated and displayed automatically. If
the configured limit of the flow rate is reached, other parameters
are also limited.
To abort the program, switch pump off.
Button
Action
ON/OFF
Switch pump on
MODE
Change to intermittent mode
Upper row +/–
Switch between the screens: dose setting (doS),
interval setting (int) and delay start (dS)