46
7.3.2
Intermediate flow rate : 125 ml/h
Sampling time : 30 seconds
Figure 7.7:
Start-up and instantenous at intermediate flow rate (125ml/h, over first 2h of the test
period)
Sampling time : 30 seconds
Figure 7.8:
Trumpet curves For 2, 5, 11, 19, 31 minutes observation windows (125 ml/h over second
hour of the test period)
Sampling time : 30 seconds
Figure 7.9:
Instantenous rate (125 ml/h, over last 2h of set change interval,24h)