User Test and Alarms checking
The following tests allow the user to check if the device is functioning
correctly.
Fresenius Kabi
recommends that these tests are performed
before attaching the
Ambix
activ
to a patient. Auditory alarm can be
checked during Downstream Occlusion detection function test.
Function
Action
Symbol
Passed
General
aspect
Check the external appearance of the pump/holder for the
absence of cracks or other visible damage.
Check that any parts seem to move freely inside the
pump/holder.
Check for the absence of visible damage on the power cord
inlet and the power cord.
Switch ON the pump, check if all display segments are
correctly displayed and check the device visually for damage
and listen for abnormal sounds.
Nurse call
cable
Check the nurse call cable is not altered or damaged (housing
and connectors). To check the proper operation of the nurse
call, create specific alarms on the pump and check that they
are replicated on the nurse call system.
Power
indication
Connect the holder with attached pump to the wall outlet.
Verify that the two-prongs plug symbol appears in the display
and the green LED on the holder lights up.
Battery
Charge
After complete charging, check the battery level indicator.
Incorrectly
installed set
Start infusion without installed set. Check that the alarm is set
off.
Empty bag
detection
Install an empty set.
Start infusion and check for empty bag alarm.
Free-flow
Check
activ
clamp visually for damage.
Connect the set to the container, open the
activ
clamp and
check flow is stopped when the clamp is closed.
Air infusion
Check the particulate/air removal filter visually for damage.
Check the proper operation of the particulate/air removal filter
of the set during priming.
Downstream
Occlusion
detection
Fill the set with the infusion solution and prime.
Occlude the downstream tubing e.g. by pinching or bending
the tubing. Start the infusion.
Check the pressure level in the display. The pressure should
increase to within the alarm zone and auditory and visual
alarms should be activated.