41
Upstream
Occlusion
detection
Administer at least 25 ml at a flow rate of 150 ml/h.
Bend the upstream tubing 2 cm above the pump.
Check if the upstream occlusion alarm is activated.
Door opened
Open the door during infusion.
Check that open door alarm is activated.
Air alarm
active
Check that the symbol
is not displayed.
Reset pump if the symbol is displayed.
Warning
In case of test failure or any doubt regarding the integrity of the
pump, please do not use the device.
If one or more checks do not comply with the right pump
behaviour, please contact the appropriate department or
Fresenius Kabi
After-Sales Service for additional verification.
Function
Action
Symbol
Passed