21
3.4.6
Start pumping
3.4.7
Stop pumping
3.4.8
Stop a program
3.4.9
Information button
3.4.10
Silence an alarm
Press the Start button to start administration.
Warning
For sets with a drip chamber, please check that the roller clamp is
opened and that liquid is flowing into the drip chamber.
Press the Stop/Esc button to stop administration.
Information
The Stop/Esc button is also used to clear alarms. In some
screens, the Stop/Esc button will also function as an escape
button to escape from an unwanted display or parameter setting
procedure.
Once a program is started, it is impossible to change values. To abort a
program, the pump has to be switched off. After switching the pump on
again, a new program can be set and started.
By pressing the i/OK button the total volume infused since the last reset
will be displayed for 5 seconds (
will be shown in the display). If the
pump is running and the target volume is activated, the remaining
volume will then be displayed (
will be shown in the display).
When the pump is not pumping, pressing the i/OK button for more than
3 seconds resets the display of the administered volume to zero.
Information
Further functions of the info button are described in the relevant
sections below
.
Alarms can be silenced for 2 minutes by pressing the Alarm silence
button (
). During this period the alarm reminder will only be indicated
by flashing of the display. It is possible to re-start administration by
pressing the Start button.