19
3.3.2
Removing the set
3.4
Operating the Ambix
activ
3.4.1
Switch pump on
3.4.2
Switch pump off
If the pump is running, stop the infusion by pressing the STOP/ESC
button.
1.
Switch the pump OFF by holding the On/Off button down for more
than 3 seconds.
2.
Ensure that backflow from the patient access system is prevented, if
present, close the roller clamp, and disconnect the set from the
patient.
3.
Open the pump door – the
activ
clamp will be automatically closed.
Remove the set starting with the green adapter.
4.
Dispose of the set following appropriate legislation and local
procedures.
5.
If it is required to restart the pump, repeat the installation procedure
described above in Section 3.3.1, page 18.
Press the On/Off button for more than one second to switch the pump
on. The pump will boot and will perform a self test. The most recent
parameters and program will be shown in the display.
Press the On/Off button for more than three seconds to switch the pump
off. The display will count down, 3-2-1-off, and the pump will then shut
down. Parameters from the most recent infusion will be retained in the
pump’s memory. This information is saved for 1 month following the last
switch-off, provided the battery was fully charged at switch off.
This operation is possible when the pump is not running (for example,
after the end of infusion programmed or at stopping of the infusion by
pressing the STOP/ESC button).