![Fresenius Kabi Ambix activ Скачать руководство пользователя страница 7](http://html.mh-extra.com/html/fresenius-kabi/ambix-activ/ambix-activ_instructions-for-use-manual_2331619007.webp)
7
1.5
Intended conditions of use
1.6
Administration Routes
1.7
Contraindications
1.8
Environment Considerations
The
Ambix
activ
is an ambulatory infusion system intended to be used
in a home environment (homecare, nursing home).
The
Ambix
activ
is made for long term use as well as over-night use.
The
Ambix
activ
is intented to be used in a backpack as an option (See
Section 11, page 55).
The nurse call function is available through the use of Ambix holder as
an option.
Outdoor use is not prohibited but temperature and humidity conditions
of use to be respected are defined:
Temperature: Operating: +13 to +40°C
Relative humidity: Operating: 20 to 85%, no condensation
Atmospheric pressure: Operating: 700 to 1060 hPa
Altitude maximum: 3000 m
The pumps administers products through intravenous (IV) access.
Not for use with life-sustaining critical application drugs, or drugs
with short half-life.
Not designed for infusion of insulin.
Not for use at flow rates below 10 ml/h.
Not suitable for neonates.
Warning
Keep the pump, sets, holder and cables away from unsupervised
children (and animals).
Consider the following operational conditions to ensure proper device
performance:
Do not expose to sun light, keep in dry place, at room
temperature, normal pressure.
Keep in clean environment.
Keep away from objects which can potentially damage the
device.
Keep away from any noise disturbance which could prevent
patient or relatives from hearing the pump alarms.
Keep away from heat source, dust, fluff, direct and prolonged
light exposure
Keep away from animals, pests or children.
Do not share an outlet with another electrical device.