30
3.6.5
Pressure sensitivity for downstream occlusion
detection
3.6.6
Alarm sound level
Button
Action
Upper +/–
Select 2 from the menu to enable KVO rate change
Lower +/–
Change to the desired value (between 0 and 10 ml/h,
increments of 1 ml/h, preset 2 ml/h)
i/OK
Confirm the adjusted value
Information
A setting of 0 will deactivate the KVO
KVO rate must be appropriate for the patient.
Pressure sensitivity settings are HI and LO. This corresponds with a
pressure level of approx. 0.8 bar and 0.5 bar, respectively.
The HI level is from 0.7 bar to 1.8 bar.
Information
During pumping, a display graphic will indicate the pressure (See
page 12).
Button
Action
Upper +/–
Select 3 from the menu to enable pressure sensitivity
change
Lower +/–
Change to the desired value (HI - LO, preset HI)
i/OK
Confirm the adjusted value
Alarm sound level settings are HI (High); ME (Medium) and LO (Low).
When an alarm is triggered, the configured value will be indicated as a
pictogram in the display. High [
], medium [
], low [ ].
Information
The time between two sound sequences can only be adjusted for
the medium and low levels (see Section 3.6.9, page 32).
Button
Action
Upper +/–
Select 4 from the menu to enable alarm volume level
change
Lower +/–
Change to the desired value (HI, ME, LO). Preset is HI.