7.2
Occlusion alarm
* Data as evaluated by the test methods of IEC 60601-2-24.
Note: the bolus volume after occlusion release (2 ml) is not considered as a hazardous situation when
the device is used for its intended use.
Warning
The internal battery should be changed every 2 years. In case of
a prolonged storage period (more than 4 months), it is
recommended that the battery is removed.
Battery replacement must be performed by qualified and trained
technical personnel in compliance with the technical manual and
procedures.
Table 7.2 - 1: Maximum time for activation of occlusion alarm at different
flow rates and Bolus volume generated when pump operating at 125
ml/h and reaching the minimum and maximum occlusion alarm
threshold*:
Occlusion alarm treshold
LO (Low)
HI (High)
Flow rate
Time of detection
Volume after
occlusion
release
Time of
detection
Volume after
occlusion
release
10 ml/h
< 10 min
< 2.0 ml
< 15 min
< 2.0 ml
125 ml/h
< 30 s
< 1.5 ml
< 1 min
< 1.5 ml