Disinfection instructions
Didecyldimethylammonium chloride shall be used as the disinfectant
agent (example : WIP'ANIOS EXCEL by Anios).
Prerequisites:
The cleaning protocol has been performed.
The pump is powered off.
The power cord and all other cables are unplugged.
The pump is disconnected from the holder.
The air is at room temperature (20 to 25 °C).
The operator is wearing suitable protective equipment.
Protocol:
1.
Place the previously cleaned pump and holder on a clean surface or
disposable underlay. The door can be removed from the pump to
facilitate the disinfection.
2.
During disinfection, do not turn the pump upside-down to avoid
liquid leaking into the pump.
3.
Use a ready-to-use wipe to wipe down all exposed surfaces of the
pump, holder and pump door (as described in cleaning protocol),
making sure to cover all cracks, crevices, and hard-to-reach areas.
Do not allow liquids to run, leak, or drip into the pump housing.
4.
Using a fresh ready-to-use wipe, repeat steps 3. Ensure that the
minimum contact time for each step is 3 minutes for bactericide
activity (surface remain visibly wet for 3 minutes). Respect the
indicated contact time from the manufacturer recommendations for
the required antimicrobial activity.
5.
Wipe down the power cord
6.
Allow the pump to dry completely at room temperature
7.
Make sure to use the original door when replacing it on the pump.