10
Cleaning and Disinfecting
10.1
Cleaning Instructions
Warning
Do not submerge the pump in water or any liquid!
Not for use in the dishwasher.
Didecyldimethylammonium chloride shall be used as the cleaning agent
(example : WIP'ANIOS EXCEL by Anios).
Prerequisites:
The pump is powered off.
The power cord and all other cables are unplugged.
The pump is disconnected from the holder.
The room temperature is between 20 and 25 °C.
The operator is wearing suitable protective equipment.
Protocol:
1.
Place the pump and the holder on a clean surface or disposable
underlay. The door can be removed from the pump to facilitate the
cleaning.
2.
During cleaning, do not turn the pump upside-down to avoid liquid
leaking into the pump.
3.
Use a ready-to-use wipe to remove any major grime.
4.
Thoroughly wipe down all exposed surfaces (housing, keyboard,
holder connection area, etc.) of the pump, from top to bottom.
Gently wipe down the exposed mechanism of the pump and the
sensor area (tube guide, clamp insert).
A cleaning process of 1 minute minimum is recommended (allow to
remain visibly wet for 1 minute), until all organic matter is dissolved
and removed.
Do not allow liquids to run, leak, or drip into the pump housing. Use
cotton wool to clean the contact pins, and hard-to-reach areas.
5.
Repeat step 4 with the pump door (housing, lever, counter door) and
holder (pole clamp screw, housing, screw area, etc.)
6.
Using a fresh ready-to-use wipe, thoroughly wipe down all exposed
surfaces. A cleaning process of 1 minute minimum is recommended
(allow to remain visibly wet for 1 minute), until all organic matter is
dissolved and removed.
7.
Wipe down the power cord
8.
Allow the device to dry completely at room temperature.
9.
Make sure to use the original door when replacing it on the pump.