Use of the internal battery
This device is provided with a Li-ion battery. When the device is disconnected from the mains, it automatically
switches to battery mode.
Before starting for the first time, charge the battery for approx. 5 hours by connecting the power supply cord
without using the device.
If the device is not used during an extended period (longer than 2 months), it is recommended to remove the
battery from the device and store it as indicated in the storage instructions. If it is not possible to remove the
battery or during a short period (less than 2 months), it is recommended to charge the battery at least once a
month, by leaving the device connected to the mains for at least 8 hours (device off).
You can also recharge a Lithium-ion battery whenever convenient, without observing the full charge/discharge
cycle that is required to get full capacity when fully charged.
In order to maximise battery lifetime and performance:
- Use and store in a cool place.
- During operation, leave the device connected to the mains to maintain the charge of the battery and the
maximum capacity when possible.
Lithium-ion rechargeable battery - to be handled with care!
- Do not incinerate or place near an open flame.
- Do not drop, crush, puncture, modify or disassemble the battery.
- Do not use a battery that is severely scarred or deformed.
- Do not short terminals.
- Do not expose to high temperature.
- Do not replace by a battery other than that specified by manufacturer.
- Do not charge or discharge otherwise than in the device.