51
10. Performance
Rates range
Volume to be infused (VTBI)
KVO (Keep Vein Open) rate
Activated when the volume to be infused is reached. KVO Rate: 1 ml/h or set flow rate (flow rate below 1 ml/h).
Modes
Rates range
Infusion rate (ml/h)
Macro
From 1 to 1500 ml/h,
with 1 ml/h increments
Maximum infusion rate can
be configured in Ward
option [Par 4] page 47
Micro
From 0.1 to 100 ml/h,
with 0.1 ml/h increments
Manual Bolus rate (ml/h)
Macro
From 200 to 1500 ml/h,
with 50 ml/h increments
Maximum infusion rate can
be configured in Ward
option [Par 9] page 47
Micro
From 200 to 1500 ml/h
with 50 ml/h increments
Programmed bolus and
loading dose
Macro
From 1 to 1500 ml/h
with 1 ml/h increments
The adjustment is common
to both modes. Maximum
infusion rate can be
configured in Ward option
[Par 9] page 47.
Micro
From 0.1 to 1500 ml/h,
by increments of 0.1 ml/h from 0.1 to
100 ml/h and 1 ml/h from 100 to 1500
ml/h
Infusion rate for a
Secondary infusion
Macro
From 1 to 1000 ml/h, with 1 ml/h
increments
Maximum infusion rate can
be configured in ward option
[Par 4], page 47.
Micro
From 0.1 to 1000 ml/h, with 0.1 ml/h
increments from 0.1 to 99.9 ml/h and
1ml/h increments from 100 to 1000 ml/h
Special case: Minimum
plateau flow rate in Ramp
mode
Macro
2 ml/h
Non-adjustable
Micro
2 ml/h
Prime rate (ml/h)
All
modes
Maximum rate (1500 ml/h)
Modes
Volume range
Volume to be infused in ml
and dose rate mode
Macro
From 1 to 9999 ml,
with 1 ml increments
Micro
From 0.1 to 1000 ml,
by increments of 0.1 ml from 0.1 to 99.9 ml and 1 ml from
100 to 1000 ml
Volume to be infused for a
secondary infusion
Macro
From 1 to 2000 ml with 1 ml increments
(Maximum value can be configured with a computer)
Micro
From 0.1 to 1000 ml with 0.1 ml increments from 0.1 to 99.9 ml and
1 ml increments from 100 to 1000 ml
(Maximum value can be configured with a computer)
Programmed bolus and
loading dose
Macro
From 1 to 100 ml,
with 1 ml increments
Micro
From 0.1 to 50 ml,
by increments of 0.1 ml
Содержание Agilia Series
Страница 1: ...Volumat MC Agilia Volumetric Infusion Pump Instructions for Use...
Страница 69: ...69 Notes...