57
Dimensions - Weight
Trumpet curves
Trumpet curves demonstrate the evolution of the minimum and maximum variance of the pump/set combination
versus flow rate.
The test protocol used to obtain these results is described in EN/IEC 60601-2-24. For further information, please
refer to this publication.
Use these curves to determine the accuracy depending upon your infusion protocol/drug/dilution. These graphs
are representative for a VL Volumat infusion set.
H/W/D
135 x 190 x 170 mm
Weight
approximately 2 kg
Screen size
70 x 35 mm
Trumpet curves for 2, 5, 11, 19, 31 minutes
observation windows
Flow rate/Time curves: start-up and instantaneous flow
rate (volume is measured every 30 seconds)
1 ml/h
1 ml/h
25 ml/h
25 ml/h
Observation windows (minutes)
% variance from nominal flow rate
-50
-40
-30
-20
-10
0
10
20
30
40
50
Accuracy upper limit (+5%)
Accuracy lower limit (-5%)
2
5
11
19
31
Accuracy upper limit (+5%)
Accuracy lower limit (-5%)
Time (minutes)
Flow rate (ml/h)
-0,2
0
0,2
0,4
0,6
0,8
1
1,2
1,4
1,6
1,8
2
2,2
2,4
2,6
2,8
3
0
20
40
60
80
100
120
Programmed flow rate
Observation windows (minutes)
% variance from nominal flow rate
-15
-10
-5
0
5
10
15
Epma
Epmin
Accuracy upper limit (+5%)
Accuracy lower limit (-5%)
2
5
11
19
31
Time (minutes)
Flow rate (ml/h)
-5
0
5
10
15
20
25
30
35
40
45
50
0
20
40
60
80
100
120
Programmed flow rate
Содержание Agilia Series
Страница 1: ...Volumat MC Agilia Volumetric Infusion Pump Instructions for Use...
Страница 69: ...69 Notes...