Warning function
Function activation
Function deactivation
Press the
key to access the menu and
select
. Press
enter.
Select an offset period to set an alarm activation
time.
Note 1:
The activation time is calculated according to
the device time, which is indicated at the bottom of the
screen.
Note 2:
If the device has been switched off during the
warning time, a warning message is displayed when
switching on the device.
Press the
key to access the menu and
select
. Press
enter.
To deactivate the function, set the offset period to
OFF.
Drop sensor (optional)
Note
: The drop sensor allows you to work in
Simple rate
infusion mode (for more details, see page 15 and
page 25) and to detect the container end. Using a drop sensor is recommended if the actual volume of the
container (bag or bottle) is not known accurately.
1. Connect the drop sensor plug to the connection socket on the back of the pump
before
switching on the
pump.
2. Fix the drop sensor to the upper part of the drip chamber by aligning the vertical part of the drop sensor with
the air inlet of the drip chamber room as indicated on the photograph.
When the drop sensor is connected, check that the symbol is displayed on the screen.
Caution
: Check the right positioning of the drip chamber and check there are no drops on the drip chamber
walls. Check that the drop sensor and the drip chamber are in a vertical position.
Note
: With the drop sensor, the maximum flow rate is restricted to 1100 ml/h.