Secondary mode
This infusion mode deliver
s
the content of a secondary bag / bottle, by means of a secondary line
connected to the main line called primary line. Once the secondary infusion is completed, the user can either
continue another secondary infusion, or resume the primary infusion (initial infusion) which parameters are
restored.
Note
: this mode is only accessible in ml/h rate mode (except : ramp, sequential and drops/min modes).
Press the
key to select the secondary mode.
Close the primary line.
Press
OK
to confirm the parameters of the primary mode and access the secondary mode.
1 - Define the infusion
2 - Secondary infusion start
In V/R mode, use the arrows to select the
secondary volume to be infused. Confirm by pressing
OK
.
Use the arrows to select the secondary flow rate.
Confirm by pressing
OK
.
Note
: the current volume infused becomes the volume
infused during the secondary infusion. It is displayed in
bold at the bottom of the screen.
Note
: the volume infused during the primary infusion is
displayed above the current volume infused.
Check the primary line is closed, check the
secondary line is connected and opened. Press
Start
to start the infusion or
C
to modify the Volume/ Rate
selection