50
9. User test
This protocol allows a quick check of pump functionality.
Volumat MC Agilia
serial number (ID/N):
_____________________________
Name:
________________________________
Ward:
________________________________
Date:
________________________________
Actions
YES
_
NO
1
. Check the state of the device:
- absence of impact marks and noises (turn the device upside down),
- presence of all labels as well as their legibility, mains lead.
2
. Without connecting the device to the mains, press the
key:
- check the functioning of the display and luminous indicators,
- functioning on the battery is signaled.
3
. Install set without liquid:
- close the door and check message: "air bubble".
4
. Remove set. Fill set with liquid. Install set incorrectly with pumping segment outside
the pump.
- check message: "Install set".
5
. Re-install set correctly as described in the user guide:
- check that the OCS test is OK.
6
. Set infusion parameters - 500 ml/h (no patient connected)
- start infusion.
7
. Clamp the upstream line with the roller clamp:
- check that the upstream occlusion alarm occurs in less than 15 seconds,
- check visual and audible alarm
8
. Open roller clamp.
9
. Start infusion (500 ml/h) and clamp the downstream line:
- check occlusion alarm (less than 15 seconds).
10
. Unclamp the downstream line. Open the door:
- check that there are no more than 3 drops falling in the drip chamber.
11
. Connect the device to the mains supply:
- check the "mains" indicator.
The device is operational when all controls are OK.
Signature