11. Technical characteristics
Power supply
Use the mains lead supplied with Volumat MC Agilia.
Battery
Disconnect battery before opening device. Avoid short circuit and excessive temperature.
If the device is not used over an extended period, all of its parameters are stored permanently, except the date
that is erased after 3 months. When the pump is switched on, you are invited to set the date again.
Communication port
The connector located at the back of the device allows different functions using the communication, mains power
and nurse call cables.
Infrared communication
Volumat MC Agilia is equipped with an infrared cell located at the back of the device. It is used for data
communication with the Agilia link rack. Data can then be transmitted by dedicated communication cables.
Compliance
Mains
Mains supply
100 V - 240 V ~/50-60 Hz with functional earth
Maximum consumption
15 VA
Protective fuses
2 x 1AT accessible in the battery compartment
External power
supply
9 VDC
/ Power > 15 W
Via a specific Fresenius Kabi accessory connected to an 8-pin connector.
Characteristics
7.4 V 2.2 Ah - Li-ion battery
Weight
Approximately 100 g
Battery life
8 h min. at the intermediate rate of 25 ml/h, and at any rate lower than 125 ml/h and
more in night mode
Battery recharge
Pump OFF: < 6 h
Pump ON: < 20 h
Nurse call
Nurse call relay output command
Serial cable
TTL output
External power
9 VDC/15 W input
Power output
5 VDC/150 mA to power Nurse Call or Serial Link accessories
Conform to the 93/42/CE Medical
Directive
IP22
Protection against splashing liquid
Protection against leakage current:
Defibrillation-proof type CF applied part
Protection against electric shocks:
class II
Functional earth
Safety of
ElectroMedical
Equipments
Compliant with EN/IEC 60601-1
and EN/IEC 60601-2-24 standards
EMC (ElectroMagnetic
Compatibility)
Compliant with EN/IEC 60601-1-2
and EN/IEC 60601-2-24 standards
The functional earth is directly connected to the mains socket. It reduces residual
current that may disturb ECG or EEG devices