Dose range
Infusion time
Drug library
Up to four drug libraries can be stored in the device. Each one can contain up to 240 drugs. This total capacity
can be limited by the number of comments and the other fields. These libraries are configured with the Vigilant
®
Drug’Lib software.
Air detection
Default setting: 250 µl detected as a single bubble or cumulated volume air over a period of 15 minutes, from
bubble sizes above 50 µl.
Resolution of sensor: ~ 10 µl.
Set replacement interval
The mechanical properties of the set in association with the pump are designed to maintain pumping
performances for 10 L maximum within a time limit of 96 hours. Nevertheless, we recommend replacing the
administration set every 24 hours for microbiological reasons, unless local policy or regulation may be
applicable. The set should be disconnected from the IV site according to local facility procedures.
Setting range
Patient data
Weight (kg)
Increment of 0.01 from 0.25 to 0.99
Increment of 0.1 from 1 to 9.9
Increment of 1 from 10 to 250
Surface (m
2
)
Increment of 0.01 from 0.05 to 4.5
Height (cm)
Increment of 1 from 20 to 250
Dilution (Units/X ml)
X ml
Increment of 1 from 1 to 9999 (Maximum value can be configured with a computer)
Units
Increment of 0.01 from 0.01 to 9.99
Increment of 0.1 from 10 to 99.9
Increment of 1 from 100 to 9999
Dose rate
Increment of 0.01 from 0.01 to 9.99
Increment of 0.1 from 10 to 99.9
Increment of 1 from 100 to 9999
Infusion time range
Loading dose
From 00 min 01 to 59 min 59
Programmed bolus
No limit. Beyond 1h, the display no longer indicates the exact time
but '>1h'
Ramp mode
From 0h01 to 48h00
Other programming modes
From 0h01 to 168h00