41
Remarks:
When a value is selected, it must be confirmed. If this value is not confirmed, the value will flash three
seconds after selection and an audible alarm is activated.
In case of a single fault condition, an alarm is activated within the limit of ± 5% rate deviation. In addition,
a secondary control feature activates an alarm at 1 ml over infusion or ±20% rate deviation, whichever is
shortest. When using a drop sensor, the flow rate is controlled at -50%, +100%.
In case of a malfunction alarm, note the error message (ErXX). Close the roller clamp, disconnect from the
mains and stop the device by pressing the
key (10 to 15 seconds can be necessary). If the alarm persists
when the device is switched on again, without use on patient, contact the qualified technicians in your
establishment or our After-Sales Department.
Drop
sensor
Connect drop
sensor
Alarm
YES
Only if drop sensor is compulsory (see page 48). At
start-up: drop sensor not connected
)
Connect the drop sensor to the pump and the
drip chamber (see page 35).
No drop sensor
Alarm
YES
Only if drop sensor is compulsory (see page 48).
During infusion or stop mode: drop sensor not
connected
)
Connect the drop sensor to the pump and the
drip chamber (see page 35).
Underflow
Alarm
YES
Flow rate detected by drop sensor is below the set
flow rate.
)
Check container. Check roller clamp. Check that
the liquid actually forms ~ 20 drops/ml and that the
drip chamber is maintained in a vertical position.
Check that the drop sensor is fixed as indicated
in page 35.
Overflow
Alarm
YES
Flow rate detected by drop sensor is above the set
flow rate.
)
Open the door and check the set positioning.
Check drop sensor positioning. Check the fluid
temperature. Check that the fluid drip forms
~ 20 drops/ml.
Uncontrolled flow
Alarm
YES
At start-up or stop mode: free flow detected by drop
sensor.
)
Close roller clamp. Check drop sensor and set
installation.
Vigilant
®
Drug’Lib
High flow rate
Warning
NO
Upper soft limit exceeded according to drug
parameters defined in drug library.
Low flow rate
Warning
NO
Lower soft limit exceeded according to drug
parameters defined in drug library.
Technical
error
Er - message
(Er01, Er02, etc.)
Alarm
YES
Technical alarm.
)
Contact your qualified technician or our
after-sales department.
System
Message
Type
Infusion
stop
Activation /
)
Action