60
Electromagnetic immunity -
Table 204
Volumat MC Agilia is intended for use in the electromagnetic environment specified below. The user of Volumat
MC Agilia should make sure it is used in such an environment.
Note 1
: At 80 MHz and 800 MHz, the highest frequency range applies.
Note 2
: These guidelines may not apply to all situations. Absorption and reflection from structures, objects and
people affect electromagnetic propagation.
(a)Field strengths from fixed transmitters, such as base stations for radio (cellular / cordless) telephones and
land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted
theoretically with accuracy. To access the electromagnetic environment due to the fixed RF transmitters, an
electromagnetic site survey should be considered. If the measured field strength in the location where
Volumat MC Agilia is used exceeds the applicable RF compliance level above, Volumat MC Agilia should
be observed to verify normal operation. If abnormal performance is observed, additional measures may be
necessary, such as re-orienting or re-locating Volumat MC Agilia, or install magnetic shielding.
(b)Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
Immunity test
IEC 60601-1-2
IEC 60601-2-24
Test level
Compliance level
obtained by the
device
Electromagnetic environment - guidance
Conducted RF
IEC 61000-4-6
10 Vrms
150 kHz to 80
MHz
10 Vrms
Portable and mobile RF communications equipment should be
used no closer to any part of the Volumat MC Agilia including
cables, than the recommended separation distance calculated
from the equation applicable to the frequency of transmitter.
Recommended separation distance:
D = 0.35
√
P , for a frequency of 150 kHz to 80 MHz
D = 0.35
√
P , for a frequency of 80 MHz to 800 MHz
D = 0.7
√
P , for a frequency of 800 MHz to 2.5 GHz
Where P is the maximum output power rating of the transmitter
in Watts (W) according to the transmitter manufacturer and
D
is the recommended separation distance in metre (m).
Radiated RF
IEC 61000-4-3
10 V/m
80 MHz to 2,5
GHz
10 V/m
Field strengths from fixed RF transmitters, as
determined by an electromagnetic site survey
(a)
,
should be less than compliance level
(b)
.
Interference may occur in the vicinity of equipment
marked with the following symbol: