Accuracy
Programmable pause
Pressure management
Occlusion alarm response time
Nominal flow rate
accuracy
± 5% on 96h with an infusion
of 10 lit
re
s maximum
In accordance with the
NF EN/IEC 60601-2-24 standard.
Manual bolus
± 5% or ± 0.2 ml
Accuracy with back
pressure of ±13.33 kPa
± 5% on 96h with an infusion of 10 lit
re
s
maximum
Programmable pause
From 1 minute to 24 h
Increments: 1 minute.
Variable mode
Maximum
pressure (B)
From 50 to 750 mmHg
25mmHg increment (50-250 mmHg)
50mmHg increment (250-750 mmHg).
Defines the authorised maximum
pressure during infusion.
Pre-alarm
level (A)
50 mmHg below
maximum pressure
Note
: If max. pressure is set to
50 mmHg, then the pre-alarm is not
activated.
3 levels mode
High
750 mmHg
These values are given as examples
and can be configurable in the User
option [User 4], page 46.
Middle
400 mmHg
Low
100 mmHg
DPS
(Dynamic Pressure
System)
Pressure
increase
Anticipates an occlusion during infusion.
Pressure
drop
A pressure decrease may indicate a disconnection or a leak in
the line.
Accuracy: the accuracy on the pressure threshold activation is 75 mmHg
or
±
15%.
Note
: 1 bar = 750 mmHg = 1000 hPa.
Rate
Occlusion alarm threshold
100 mmHg
300 mmHg
400 mmHg
750 mmHg
1 ml/h
7’
22’
34’
58’
25 ml/h
15"
45"
56"
1’ 52"
100 ml/h
2"
9"
12"
27"
These values can vary by
±
20% depending on the device and the infusion set.