40
Pressure
Pressure increase
(DPS)
Warning
NO
The pressure is increasing in the line. This warning
can be selected as an option [User 4] see page 46.
)
Check if infusion line is occluded (stop-cock,
catheter, folded line).
Occlusion
pre-alarm
Pre-
alarm
NO
In-line pressure has reached 50 mmHg below the
programmed threshold.
)
Check the infusion line.
Set the correct pressure threshold.
Downstream
occlusion
Alarm
YES
The pressure in the line has reached the threshold
level (see page 31).
)
Check if the infusion line is occluded (stop-cock,
catheter, folded line). If necessary, readjust
pressure threshold in relation to flow rate.
Upstream
occlusion
Alarm
YES
Only without drop sensor. The pressure in the
upstream line is too low.
)
Check the roller clamp. Check the container
and line. Check the container height. Check air
inlet cap (if a bottle is used) folded line.
Pressure drop
(DPS)
Warning
NO
Pressure drop in infusion line. This warning can be
selected as an option.
)
Check the downstream Luer Lock connection
and full line integrity.
Battery
Battery pre-alarm
Pre-
alarm
NO
Low battery.
)
Connect to mains supply.
Battery alarm
Alarm
YES
Discharged battery.
)
The pump will turn OFF automatically within 5
minutes. Connect the pump to the mains supply
immediately.
Empty battery
Alarm
YES
Connect to mains supply and wait for battery to be
charged.
Mains
Mains
disconnection
Warning
NO
Mains supply disconnection.
)
Press silence to acknowledge and check if
battery life is sufficient for the expected infusion
duration. If the disconnection is unintentional,
check the mains connection.
System
Message
Type
Infusion
stop
Activation /
)
Action