59
12. Guidance and manufacturer’s
declaration on EMC
Electromagnetic emissions -
Table 201
Volumat MC Agilia is intended for use in the electromagnetic environment specified below. The user of Volumat
MC Agilia should make sure it is used in such an environment.
Electromagnetic immunity -
Table 202
Volumat MC Agilia is intended for use in the electromagnetic environment specified below. The user of Volumat MC Agilia should
make sure it is used in such environment.
Emissions test
Compliance
obtained by the
device
Electromagnetic environment - guidance
RF emissions
CISPR 11
Group 1
Volumat MC Agilia uses RF energy only for its internal function. Therefore, its RF emissions are
very low and are not likely to cause any interference in nearby electronic equipment.
RF emissions
CISPR 11
Class B
Volumat MC Agilia is suitable for use in all establishments, including domestic and hospital
establishments and those directly connected to the public low-voltage power supply network that
supplies buildings used for domestic purposes.
Harmonic emissions
IEC 61000-3-2
Class A
Voltage fluctuations
Flicker emissions
IEC 61000-3-3
Does not apply
Immunity test
IEC 60601-1-2
IEC 60601-2-24
Test level
Compliance level
obtained by the
device
Electromagnetic environment - guidance
Electrostatic
Discharge (ESD)
IEC 61000-4-2
±
8 kV contact
±
15 kV air
±
8 kV contact
±
15 kV air
Coatings of the floors out of wooden, tilling, and concrete, with a relative
humidity level at least 30 %, make it possible to guarantee the level of
necessary conformity. If it is not possible to guarantee this environment,
additional precautions must be taken, such as: anti-static material usage,
preliminary user discharge and the wearing of anti-static clothing.
Electrical fast
Transient / burst
IEC 61000-4-4
±
2 kV for power
supply lines
±
1 kV
for input output lines
±
2 kV for power
supply lines
±
1 kV
for input output lines
Mains power quality should be that of a typical domestic, commercial or
hospital environment.
Surge
IEC 61000-4-5
±
1 kV
differential mode
±
2 kV
common mode
±
1 kV
differential mode
±
2 kV
common mode
Mains power quality should be that of a typical domestic, commercial or
hospital environment.
Voltage dips, short
interruptions and
voltage variations on
power supply input
lines
IEC 61000-4-11
< 5 % Ut
(> 95 % dip in Ut)
for 0,5 cycle
< 5 % Ut
(> 95 % dip in Ut)
for 0,5 cycle
Mains power quality should be that of a typical domestic, commercial or
hospital environment.
For short and long interruptions (< than battery life) of power mains, the
internal battery provides the continuity of service.
Note
: Ut is the a/c. main voltage prior to application of the test level.
40 % Ut
(60 % dip in Ut)
for 5 cycles
40 % Ut
(60 % dip in Ut)
for 5 cycles
70 % Ut
(30 % dip in Ut)
for 25 cycles
70 % Ut
(30 % dip in Ut)
for 25 cycles
< 5 % Ut
(> 95 % dip in Ut)
for 5 s
< 5 % Ut
(> 95 % dip in Ut)
for 5 s
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
400 A / m
400 A / m
If necessary, the power magnetic field should be measured in the
intended installation location to assure that it is lower than compliance
level.
If the measured field in the location where the Volumat MC Agilia is used
exceeds the applicable magnetic field compliance level above, the
Volumat MC Agilia should be observed to verify normal operation. If
abnormal performance is observed, additional measures may be
necessary, such as re-orienting or re-locating Volumat MC Agilia, or
install magnetic shielding.