13. Cleaning and use conditions
Cleaning and disinfecting
Volumat MC Agilia is part of the patient’s immediate environment. It is advisable to clean and disinfect the
device external surfaces regularly and especially before connecting a new patient and before any maintenance
operation in order to protect patient and staff.
1. Prepare the detergent-disinfectant solution.
2. Disconnect the device from the power supply.
3. Moisten the disposable cloth with the detergent-disinfectant solution, carefully wring out the cloth. Repeat at
each stage of the cleaning process.
4. Start by cleaning the bottom side of the device. Then carefully turn the device upside down without touching
the mobile parts. Put down the device on a clean surface.
5. Continue the cleaning on sides of the device without wetting the sockets.
6. Clean the keyboard.
7. Complete the cleaning of the most exposed surfaces, the most critical zones and the mains cord.
8. Do not rinse, leave to dry.
9. Protect and keep the device clean before reuse.
10.Validate the maintenance protocol by simple bacteriological checking.
Do not place in an AUTOCLAVE or IMMERSE the device. Do not let liquids enter the device’s casing.
DO NOT USE
: TRICHLOROETHYLENE-DICHLOROETHYLENE - AMMONIA - AMMONIUM CHLORIDE -
CHLORINE and AROMATIC HYDROCARBON - ETHYLENE DICHLORIDE-METHYLENE CHLORIDE -
CETONE. These aggressive agents could damage the plastic parts and cause device malfunction.
Take care also with ALCOHOL BASED SPRAYS (20% - 40% alcohol). They lead to tarnishing and create
small cracks in the plastic, and do not provide the necessary cleaning prior to disinfecting. Disinfecting SPRAYS
may be used, in accordance with the manufacturer recommendation, from a distance of 30 cm of the device,
avoid the accumulation of the product in liquid form.
Please contact the appropriate service, responsible for cleaning and disinfecting products, in your
establishment for further details.
Environmental conditions
The device should be stored in a dry and cool place. In case of prolonged storage, the battery should be
disconnected. This should be done by a qualified technician who can access the battery via the battery access
flap situated underneath the device.
Storage conditions and carrying
Temperature: - 10°C to +60°C.
Pressure
: 500 hPa to 1060 hPa.
Humidity
: 10% to 90%, no condensation
Use conditions
Temperature: 5°C to 40°C.
Pressure
: 700 hPa to 1060 hPa.
Humidity
: 20% to 90%, no condensation.