[Par 31] Pre alarm end
VTBI
Defines the pre-alarm parameters
Note
: The parameters of the pre-alarms can be readjusted.
Nevertheless, the following situation should be carefully considered:
the end of infusion pre-alarm should not be deactivated in case of
short half-life drugs or for infusion with rigid bottles.
Duration: from 0 to 30 min
before the end of the infusion
Note
: Adjusting down to 0
(disabling) is possible with
Partner Agilia software only.
% Volume: from 0 to 15% of
remaining VTBI
Volume: from 0 to 50 ml of
remaining VTBI
With drop sensor: Activate or
deactivate pre-alarm when using
drop sensor
[Par 33] Drops/min.
Volumic flow rate
Allows
selection of
adjustable flow
rates
Drops/min flow
rate
[Par 34] Rate modification
Allows modification of the flow rate in STOP mode only
Maint.
Maintenance