16
MyCrown Mill
A
Number of test runs completed since the last time the Start button was actuated.
B
Progress bar for test currently being run.
C
Check boxes for selecting (activating/deactivating) the tests available in this test dialog.
D
Check box for selecting the options possible in this test dialog.
E
Number of test results since the last start:
● Passed fully
● Passed
● Not passed
F
With the
"Start"
button, the test run is started with the settings selected. The test run is repeated until it is
cancelled using the
"Stop"
button. The test run counter is then reset to 0. After starting, the test run can only
be paused with
"Stop"
. No other inputs are possible during the test run.
G
With the
"Stop"
button, the test is broken off as quickly as possible. The test in progress at the time of
cancellation is not counted. All inputs are now possible again.
H
Store the available data in the log file with the
"Save"
button. The data are appended to the previously existing
log file (if the current test run has not yet been saved). If no data is available, a message appears to this effect
(confirm with
"OK"
). Once a test run has been saved to the log file, a new test run must be started before
selecting the save function again.
I
With
"Back"
, you can exit the test dialog and return to the service dialog. If any data exist and have not yet
been saved, you will be asked whether or not the test data should be saved.
J
Result:
● No color: No measurement available yet.
● green: Test passed fully.
● yellow: Test passed.
● red: Test not passed.
Содержание MyCrown Mill
Страница 1: ...j j p j b Service Manual...
Страница 15: ...15 Service Manual b Example Total test Total test...
Страница 33: ...33 Service Manual b 3 1 Force measurement errors...
Страница 34: ...34 MyCrown Mill 3 2 Stepping motor does not move...
Страница 35: ...35 Service Manual b 3 3 Spindle motor does not lock...
Страница 36: ...36 MyCrown Mill 3 4 Milling chamber door switch...
Страница 37: ...37 Service Manual b 3 5 Unit switches off Continued on next page...
Страница 38: ...38 MyCrown Mill Continued from previous page...
Страница 39: ...39 Service Manual b 3 6 Edges break off holes or wide stripes in restoration Continued on next page...
Страница 40: ...40 MyCrown Mill Continued from previous page...
Страница 41: ...41 Service Manual b 3 7 Insufficient air pressure...
Страница 42: ...42 MyCrown Mill 3 8 Stepping motor cannot be controlled...
Страница 43: ...43 Service Manual b 3 9 Faulty water pressure...
Страница 44: ...44 MyCrown Mill Continued from previous page...
Страница 45: ...45 Service Manual b 3 10 Unit cannot find starting position...
Страница 46: ...46 MyCrown Mill 3 11 Unit cannot be turned on...
Страница 47: ...47 Service Manual b 3 12 Touch errors...
Страница 48: ...48 MyCrown Mill 3 13 Service life of the milling and grinding machine is too short...
Страница 49: ...49 Service Manual b 3 14 Milling unit not addressable...
Страница 50: ...50 MyCrown Mill 3 15 LAN communication problems Continued on next page...
Страница 51: ...51 Service Manual b Continued from previous page...
Страница 52: ...52 MyCrown Mill 3 16 WLAN connection occasionally interrupted...
Страница 53: ...53 Service Manual b 3 17 WLAN communication problems...
Страница 67: ...67 Service Manual b 4 5 Replacing stepping motors Overview of stepping motors...