10
MyCrown Mill
2.1.
General information
Test requirements
Requirement for all tests:
● PC / acquisition unit is switched on and ready for operation.
● PC/ acquisition unit and milling unit are connected to each other
(network cable, wireless or WLAN).
● The door of the milling chamber must remain closed as long as any
motors or the water pump are running. If the door of the milling
chamber is opened during the test, all motors and the water pump will
switch off immediately (same function as pressing the
"STOP"
button).
● When operating the water pump, the tools (grinder or miller) must be
installed.
● When operating the water pump, the air pump must always be
running alongside it in order to protect the DC motors from water,
e.g., the air pump is started first. If the air pressure switch detects no
air pressure, an error message will appear and the test cannot be
performed.
● To operate the suction system, if the machine configuration permits
it, the suction system must be connected to the milling unit (interface
cable and suction hose) and switched to “
Auto
”.
Service software log file
The Service software generates a log file for all tests performed. This log
file is located in the folder on delivery.
C:\Program Files\FONA Dental\MyCrown Design
1.0.x\Bin\Service\Service\Protocols
The protocol is called:
Test file_XXXXXXXXXXXX_YYYYYYYY_ZZZZZZ.xml or
Test file_XXXXXXXXXXXX_YYYYYYYY_ZZZZZZ.html
The “X” here denotes the serial no. of the milling unit, “Y” denotes the
date, and “Z” the time the protocol was generated.
Each time the service software is started, a confirmation query appears
asking whether this file (if it already exists) should be deleted. If
"No"
(do
not delete) is selected, the tests subsequently performed will be
appended to the ones previously saved. Each test is labeled with a
starting and ending date. The test file can be viewed with the editor at any
time. If the user quits a test without saving it, he will be asked if he really
wants to quit without saving. If
"Yes"
(Quit without saving) is then
selected, the data just measured will be lost.
Assessment scores
There are three different scores which can be assigned to test results:
● Passed fully (
"Passed fully"
, green label)
● Passed (
"Passed"
, yellow label)
● Not passed (
"Not passed"
, red label)
The area which passed may be irrelevant for a specific test. In that case,
n/a (not applicable) will be written to the log file for this area.
Содержание MyCrown Mill
Страница 1: ...j j p j b Service Manual...
Страница 15: ...15 Service Manual b Example Total test Total test...
Страница 33: ...33 Service Manual b 3 1 Force measurement errors...
Страница 34: ...34 MyCrown Mill 3 2 Stepping motor does not move...
Страница 35: ...35 Service Manual b 3 3 Spindle motor does not lock...
Страница 36: ...36 MyCrown Mill 3 4 Milling chamber door switch...
Страница 37: ...37 Service Manual b 3 5 Unit switches off Continued on next page...
Страница 38: ...38 MyCrown Mill Continued from previous page...
Страница 39: ...39 Service Manual b 3 6 Edges break off holes or wide stripes in restoration Continued on next page...
Страница 40: ...40 MyCrown Mill Continued from previous page...
Страница 41: ...41 Service Manual b 3 7 Insufficient air pressure...
Страница 42: ...42 MyCrown Mill 3 8 Stepping motor cannot be controlled...
Страница 43: ...43 Service Manual b 3 9 Faulty water pressure...
Страница 44: ...44 MyCrown Mill Continued from previous page...
Страница 45: ...45 Service Manual b 3 10 Unit cannot find starting position...
Страница 46: ...46 MyCrown Mill 3 11 Unit cannot be turned on...
Страница 47: ...47 Service Manual b 3 12 Touch errors...
Страница 48: ...48 MyCrown Mill 3 13 Service life of the milling and grinding machine is too short...
Страница 49: ...49 Service Manual b 3 14 Milling unit not addressable...
Страница 50: ...50 MyCrown Mill 3 15 LAN communication problems Continued on next page...
Страница 51: ...51 Service Manual b Continued from previous page...
Страница 52: ...52 MyCrown Mill 3 16 WLAN connection occasionally interrupted...
Страница 53: ...53 Service Manual b 3 17 WLAN communication problems...
Страница 67: ...67 Service Manual b 4 5 Replacing stepping motors Overview of stepping motors...