Operating Manual CMGZ 622
66
Setting the analog output (jumper)
Channel 1
(main board)
Channel 2...4
(subprint)
Analog output
0...10V
(default)
Analog output
±10V
J1100
J100 2-3
1-2
Setting the sensor excitation (solder bridges)
Channel 1
Channel 2...4
Sensor excitation
(main board)
(subprint)
5VDC
(default)
10VDC 24VDC
J1200
J200
1-2
2-3
2-3
J1201
J201 open
closed
closed
J1202
J202 open
open
closed
J1203
J203
closed
closed
open
J1204
J204 open
open
closed
J1205
J205
closed
closed
open
Setting the sensor signal (solder bridges)
Channel 1
(main board)
Channel 2...4
(subprint)
Sensor signal
±12.5 or ±25mV
(default)
Sensor signal
0...10V
J1201
J201 open
1)
closed
1)
closed
J1208
J208
closed
open
J1209
J209
closed
open
J1210
J210 open
closed
J1211
J211 open
closed
J1212
J212
closed
open
1)
Depending on sensor excitation, see above
Setting to 4 wire or 6 wire circuit (solder bridges)
Channel 1
(main board)
Channel 2...4
(subprint)
4 wire circuit
(default)
6 wire circuit
J1206
J206
closed
open
J1207
J207
closed
open
Note
The jumpers and solder bridges are normally factory set and need no customization.