Operating Manual CMGZ 622
40
14.4
Description of the System Parameters
The parameter changing mode is activated by pressing the
PARA
↵
key for 3 seconds.
By pressing the
PARA
↵
key again, the system parameters are selected (ref. also to fig.
19).
Language
Use:
This parameter stores the display language.
Range:
English, French, Italian, German
Measuring system
Use:
This parameter indicates the measuring system to be used. If it is set
to
metric
, all force values are shown as [N, cN, kN]. If it is set to
US
standard
, all force values are shown as [lb, clb, klb].
Range:
Metric, US standard
Default:
Metric
Lowpass display
Use:
The electronic unit provides a lowpass filter to prevent noise which is
added to the integrated display. This parameter stores the cut off
frequency. The lower the cut off frequency, the more sluggish the
output signal will be. Due to this filter, the value shown in the display
will be much more stable in the case of high fluctuations of the force
value. The lowpass display filter is independent to the other filters.
Range:
0.1
to
10.0
Default:
1.0
Increment:
0.1
Unit:
[Hz]
Identifier
Use:
This parameter stores the ident number of the device when linked to
PROFIBUS, CAN-Bus resp. DeviceNet.
Range:
0
to
255
Default:
0
Increment:
1
Unit:
[-]
Baud rate
Use:
This parameter stores the speed of the serial interface (RS232).
Range:
2400, 4800, 9600, 19200
Default:
9600
Unit:
[Baud]
Note:
The other serial interface settings are fixed: 8 data bits, even parity,
1 stop bit („8 e 1“).