3. OPERATION
Page 61
© 2005-2007 DH Instruments, a Fluke Company
3.5.1.1
EDIT AUTOZ
The edit AutoZ function should be used with great caution as entering
inappropriate values and turning AutoZ ON may result in out of tolerance
measurements. In normal operation, the value of the AutoZ offset, P
offset
,
should be changed by running AutoZ using [AutoZ] (seen Section 3.3.9).
Before editing P
offset
, see Section 3.5.1,
PRINCIPLE.
To edit the current
P
offset
value for the active Q-RPT and measurement mode,
press
[SPECIAL]
,
<1AutoZ>
,
<3edit>
. The display is:
1. Indication of active measurement mode
(
<abs>
for absolute,
<gage>
for gauge and
negative gauge,
<dif>
for differential).
2.
Active Q-RPT for which
P
offset
is being edited.
3.
Edit field for value of
P
offset
.
Poffset, Abs Hi
8.4 Pa
Edit
P
offset
as desired and press
[ENT]
to activate the new value. Press
[ESC]
to
abandon changes.
If editing the HL (parallel mode)
P
offset
values, the editing display shows the
values for both Q-RPTs simultaneously:
1. Indication of measurement mode for which
this is
P
offset
(
<a>
for absolute,
<g>
for gauge).
2.
Edit field for value of Hi Q-RPT
P
offset
.
3.
Indication that this is the Hi Q-RPT
P
offset
.
4.
Edit field for value of Lo Q-RPT
P
offset
.
5.
Indication that this is the Lo Q-RPT
P
offset
.
Poff,a: 1 Pa Hi
Poff,a: -1 Pa Lo
The value of P
offset
is always in Pascal (Pa). The value of P
std,0
is entered in
the current pressure unit of measure.
3.5.2
<2REMOTE>
PURPOSE
To configure the RPM4-AD COM1, COM2 and IEEE-488 (GPIB) communication ports. To test
COM1 and COM2 communications. To select the remote programming communications format.
PRINCIPLE
The RPM4-AD has two RS-232 communications ports referred to as COM1 and COM2 and a
single IEEE-488 (GPIB) port. COM1 or the IEEE-488 port is for communicating with a host
computer (see Section 4), and COM2 is reserved for communicating with an external device
(e.g. an RPMx, a multimeter, etc.). These ports’ settings can be viewed and changed using
[SPECIAL]
,
<2remote>
.
RPM4-AD has two remote communications formats, classic and enhanced (see Section 4.3).
Which of these is active can be selected.
A self test is supplied for RS-232 communications. The self test allows verification that the
RPM4-AD RS232 ports (COM1 and COM2) are operating properly and that a valid interface
cable is being used.
Содержание RPM4-AD
Страница 10: ...RPM4 AD OPERATION AND MAINTENANCE MANUAL 2005 2007 DH Instruments a Fluke Company Page VIII N NO OT TE ES S...
Страница 22: ...RPM4 AD OPERATION AND MAINTENANCE MANUAL 2005 2007 DH Instruments a Fluke Company Page 12 N NO OT TE ES S...
Страница 90: ...RPM4 AD OPERATION AND MAINTENANCE MANUAL 2005 2007 DH Instruments a Fluke Company Page 80 N NO OT TE ES S...
Страница 138: ...RPM4 AD OPERATION AND MAINTENANCE MANUAL 2005 2007 DH Instruments a Fluke Company Page 128 N NO OT TE ES S...
Страница 142: ...RPM4 AD OPERATION AND MAINTENANCE MANUAL 2005 2007 DH Instruments a Fluke Company Page 132 N NO OT TE ES S...