3. OPERATION
Page 17
© 2005-2007 DH Instruments, a Fluke Company
AN RPM4-AD Q-RPT is identified by a range screen identifying the Q-RPT, its current unit of measure
and its full scale pressure in gauge and absolute measurement modes or in an air data unit. The
Q-RPT screen is:
1. Q-RPT
designator.
2.
Q-RPT position designator.
3.
Current pressure unit of measure.
4. Full scale pressure in current unit of measure in gauge (
<g>
) and/or absolute
(
<a>
) measurement mode.
Active A160K Hi
inHg 17.7g/47.2a
The Q-RPTs available on an RPM4-AD and short cuts to specific air data measurement
modes are accessed using
[RANGE]
.
Most settings made when an RPM4-AD Q-RPT is active, such as unit of measure,
measurement mode, display resolution, and stability setting are specific to the Q-RPT.
Generally, settings selected while one Q-RPT is active apply to that Q-RPT and not to the
other Q-RPT. The Q-RPT specific settings are stored with the Q-RPT and recalled whenever
the Q-RPT is made active. See Table 5 for a listing of RPM4-AD settings and to what level of
RPM4-AD operation they are specific.
Table 5.
Settings and what they are specific to
(range, measurement mode, Q-RPT, system)
SETTING
PURPOSE
SPECIFIC TO
SECTION
[Unit]
Set pressure unit of measure
Q-RPT and measurement mode
3.3.2
[Mode]
Set pressure measurement mode (absolute,
gauge, negative gauge)
Q-RPT
3.3.3
[Display]
Set bottom line display function
System
3.3.6
[Head]
Set fluid head correction height, fluid, unit of
measure
System
3.3.7
[SDS]
Open and close SDS
Q-RPT
3.3.8
[AutoZ]
Run AutoZ
Q-RPT and measurement mode
3.3.9
Resolution
Set pressure display resolution
Q-RPT
3.4.2
Stability
Set
Ready/Not Ready
stability test.
Q-RPT
3.4.3
Upper Limit
Set upper and lower pressure limit alarm
Q-RPT and measurement mode
3.4.4
AutoZ
AutoZ ON/OFF, set and view values
Q-RPT and measurement mode
3.5.1
Screen Saver,
Sound, Time,
ID, Level
Set system user preferences
System
3.5.5
ReadRate
Automated read rate adjustment ON/OFF
System
3.5.7.2
Cal
Q-RPT and barometer calibration coefficient
and date viewing and editing
Q-RPT or barometer
3.5.7.5
3.2.2
GAUGE AND NEGATIVE GAUGE MEASUREMENT,
DYNAMIC COMPENSATION FOR ATMOSPHERIC
PRESSURE
RPM4-AD Q-RPTs are intrinsically absolute but they are also used for gauge and negative gauge
measurement modes (see Section 3.3.3,
PRINCIPLE). Gauge measurement modes are
achieved by subtracting the value of atmospheric pressure,
P
offset,G
, from the Q-RPT’s absolute
reading using AutoZ (see Section 3.2.2). The AutoZ routine that measures
P
offset,G
, is run by
pressing
[AutoZ]
whenever RPM4-AD is in the vented condition. This assures the continuous
Содержание RPM4-AD
Страница 10: ...RPM4 AD OPERATION AND MAINTENANCE MANUAL 2005 2007 DH Instruments a Fluke Company Page VIII N NO OT TE ES S...
Страница 22: ...RPM4 AD OPERATION AND MAINTENANCE MANUAL 2005 2007 DH Instruments a Fluke Company Page 12 N NO OT TE ES S...
Страница 90: ...RPM4 AD OPERATION AND MAINTENANCE MANUAL 2005 2007 DH Instruments a Fluke Company Page 80 N NO OT TE ES S...
Страница 138: ...RPM4 AD OPERATION AND MAINTENANCE MANUAL 2005 2007 DH Instruments a Fluke Company Page 128 N NO OT TE ES S...
Страница 142: ...RPM4 AD OPERATION AND MAINTENANCE MANUAL 2005 2007 DH Instruments a Fluke Company Page 132 N NO OT TE ES S...