3. OPERATION
Page 29
© 2005-2007 DH Instruments, a Fluke Company
If the unit of measure selected is a true airspeed (unit label in upper case, e.g. KTS), the
temperature used in airspeed calculation must be specified. When the unit is selected, the
display is:
1. Entry field for temperature to be used in true airspeed
calculation.
True airspeed temp:
15.0 ºC
Enter the desired temperature value. The default value for true airspeed air temperature is
set in
[SPECIAL]
,
<6Punit>
when the unit is set up (see Section 3.5.6).
In RPM4-AD displays, calibrated airspeed units are represented in lower case (e.g. kts)
and true airspeed units are represented in upper case (e.g. KTS).
Table 9.
Default
[UNIT]
choices by active Q-RPT and measurement mode,
RPM4 A160K/A160K (rotary wing)
UNIT
Q-RPT
MEASUREMENT
MODE
#1
#2
#3
#4
#5
#6
Absolute
<a>
kPa inHg psi mbar m
ft
Hi and Lo
parallel
<HL>
Gauge,
negative gauge
<g>
kPa
inHg psi inWa kts km/h
Absolute
<a>
kPa inHg psi mbar hPa mmHg
Hi
<Hi>, <Hd>
Gauge,
negative gauge
<g>
,
differential
<d>
1
kPa
inHg psi inWa kts km/h
1. Altitude measured by Lo Q-RPT can be displayed while measuring airspeed.
Table 10.
Default
[UNIT]
choices by active Q-RPT and measurement mode,
RPM4 A350K/A160K (rotary wing)
UNIT
X
MEASUREMENT
MODE
#1
#2
#3
#4
#5
#6
Absolute
<a>
kPa inHg psi mbar m
ft
Lo
<Lo>
Gauge,
negative gauge
<g>
kPa inHg psi inWa mbar hPa
Absolute
<a>
kPa inHg psi mbar m
ft
Hi
<Hi>, <Hd>
Gauge,
negative gauge
<g>
,
differential
<d>
1
kPa
inHg psi inWa kts km/h
1. Altitude measured by Lo Q-RPT can be displayed while measuring airspeed.
The pressure unit of measure selected is Q-RPT and measurement mode specific. For that
Q-RPT and measurement mode, all functions and settings are represented in the current
unit of measure for that range. However, certain internal and/or metrological functions
(e.g., Q-RPT calibration coefficients) are always represented in Pa regardless of the current
unit selection.
See
Section
7.2 for information on the unit of measure
conversion
factors used by RPM4-AD.
Содержание RPM4-AD
Страница 10: ...RPM4 AD OPERATION AND MAINTENANCE MANUAL 2005 2007 DH Instruments a Fluke Company Page VIII N NO OT TE ES S...
Страница 22: ...RPM4 AD OPERATION AND MAINTENANCE MANUAL 2005 2007 DH Instruments a Fluke Company Page 12 N NO OT TE ES S...
Страница 90: ...RPM4 AD OPERATION AND MAINTENANCE MANUAL 2005 2007 DH Instruments a Fluke Company Page 80 N NO OT TE ES S...
Страница 138: ...RPM4 AD OPERATION AND MAINTENANCE MANUAL 2005 2007 DH Instruments a Fluke Company Page 128 N NO OT TE ES S...