3. OPERATION
Page 25
© 2005-2007 DH Instruments, a Fluke Company
OPERATION
Pressing the
[RANGE]
function key activates the Q-RPT viewing and selecting function.
When the
[RANGE]
function key is first pressed, the active Q-RPT is displayed. For example:
1. Q-RPT
designator.
2. Q-RPT
position
designator
(
<Hi>
,
<Lo>
,
<HL>
for both Q-RPTs in
parallel mode,
<Hd>
for Hi Q-RPT in differential mode with the Lo Q-RPT).
3.
Current unit of measure.
4. Full scale pressure of Q-RPT in current unit of measure in gauge (
<g>
) and
absolute (
<a>
) measurement modes for pressure unit of measure or in air
data unit.
Active A160K Hi
kPa 60g/160a
Pressing the
[RANGE]
key again while in the RANGE functions toggles through displays of
the other available Q-RPTs and shortcuts to air data measurement configurations (see below for
RPM4-AD A160K/A160K and RPM4-AD A350K/A160K
[RANGE]
key choice details). The
[
]
and
[
]
keys can also be used to move forwards or backwards through the selections.
To select a Q-RPT or shortcut, press
[ENT]
when it is displayed.
Pressing
[ESC]
while in the RANGE function returns to the MAIN RUN screen without
making any change.
Q-RPT specific RPM4-AD operating settings are saved with the Q-RPT when another Q-RPT
is made active (see Section 3.2.1, Table 5). When the Q-RPT is selected again, it has the
same settings as it was left with. The air data measurement shortcuts always return to the
Q-RPT and measurement mode necessary for the air data measurement. They retain all
other user settings.
3.3.1.1
RPM4-AD A160K/A160K (ROTARY WING)
Pressing
[RANGE]
always displays the active Q-RPT information. Pressing
[RANGE]
again toggles through the other Q-RPT choices available and the three
air data shortcuts. There are a total of five
[RANGE]
displays.
The first time
[RANGE]
is pressed, the active range is displayed. Pressing
[RANGE]
again displays the next range choice. Pressing
[ENT]
activates the
displayed selection and returns to the main run screen.
The sequence of selections when
[RANGE]
is pressed is:
FIRST [RANGE] SELECTION: HI Q-RPT
Selects Hi Q-RPT with all previous
settings (measurement mode, unit of
measure, etc) for that Q-RPT.
Select A160K Hi
kPa 60g/160a
SECOND [RANGE] SELECTION: HL (PARALLEL MODE) Q-RPT
Selects HL Q-RPT (Hi and Lo Q-RPTs in
parallel ) with all previous settings
(measurement mode, unit of measure,
etc) for that Q-RPT.
Active A160K HL
kPa 60g/160a
Содержание RPM4-AD
Страница 10: ...RPM4 AD OPERATION AND MAINTENANCE MANUAL 2005 2007 DH Instruments a Fluke Company Page VIII N NO OT TE ES S...
Страница 22: ...RPM4 AD OPERATION AND MAINTENANCE MANUAL 2005 2007 DH Instruments a Fluke Company Page 12 N NO OT TE ES S...
Страница 90: ...RPM4 AD OPERATION AND MAINTENANCE MANUAL 2005 2007 DH Instruments a Fluke Company Page 80 N NO OT TE ES S...
Страница 138: ...RPM4 AD OPERATION AND MAINTENANCE MANUAL 2005 2007 DH Instruments a Fluke Company Page 128 N NO OT TE ES S...
Страница 142: ...RPM4 AD OPERATION AND MAINTENANCE MANUAL 2005 2007 DH Instruments a Fluke Company Page 132 N NO OT TE ES S...