RPM4 Q-RPT is active and the active measurement mode. See Table 7 or Table 8 for
definition of the types of units of measure are available by Q-RPT and measurement mode.
The choice of six units displayed by the
[UNIT]
function can be customized by the user (see
Section 3.5.6).
Table 7.
Available unit of measure types by active Q-RPT and measurement mode,
RPM4-AD A160K/A160K (rotary wing)
UNIT TYPE
-RPT
MEASUREMENT
MODE
PRESSURE
ALTITUDE
AIRSPEED
USER
Absolute
<a>
Yes
1
Yes
1
No Yes
1
Hi and
Lo
parallel
<HL>
Gauge,
negative gauge
<g>
Yes
1
No
2
Yes
1,
2
Yes
1
Absolute
<a>
Yes No No
Yes
Hi
<Hi>,
<Hd>
Gauge
,
negative gauge
<g>
,
differential
<d>
Yes No
2
Yes
2
Yes
1. Best measurement uncertainty .
2. Altitude measured by Lo Q-RPT can be displayed while measuring airspeed.
Table 8.
Available unit of measure types by active Q-RPT and measurement mode,
RPM4-AD A350K/A160K (fixed wing)
UNIT TYPE
Q-RPT
MEASUREMENT
MODE
PRESSURE
ALTITUDE
AIRSPEED
USER
Absolute
<a>
Yes
Yes
No Yes
Lo
<Lo>
Gauge,
negative gauge
<g>
Yes No No
Yes
Absolute
<a>
Yes No No
Yes
Hi
<Hi>,
<Hd>
Gauge,
negative gauge
<g>
,
differential
<d>
Yes No
1
Yes
1
Yes
1. Altitude measured by Lo Q-RPT can be displayed while measuring airspeed.
Internally, RPM4-AD always operates in Pascal (Pa), the SI unit of pressure. Values
expressed in other units of measure are obtained by applying conversion or scaling factors to
Pascal. (see Section 7.2).
OPERATION
To change the active unit of measure for a Q-RPT and
measurement mode, press
[UNIT]
from the MAIN
RUN screen while the Q-RPT and measurement
mode is active (see Table 9 or Table 10 for default
units of measure by active Q-RPT and measurement
mode). The display is (actual units shown specific to
Q-RPT and measurement mode):
1kPa 2inHg 3psi
4mbar 5m 6ft
The cursor is on the number corresponding to the active pressure unit of measure for the active range.
To change the pressure unit of measure, select the desired unit. Making the selection
returns to the MAIN RUN screen with the selected unit active.
Содержание RPM4-AD
Страница 10: ...RPM4 AD OPERATION AND MAINTENANCE MANUAL 2005 2007 DH Instruments a Fluke Company Page VIII N NO OT TE ES S...
Страница 22: ...RPM4 AD OPERATION AND MAINTENANCE MANUAL 2005 2007 DH Instruments a Fluke Company Page 12 N NO OT TE ES S...
Страница 90: ...RPM4 AD OPERATION AND MAINTENANCE MANUAL 2005 2007 DH Instruments a Fluke Company Page 80 N NO OT TE ES S...
Страница 138: ...RPM4 AD OPERATION AND MAINTENANCE MANUAL 2005 2007 DH Instruments a Fluke Company Page 128 N NO OT TE ES S...
Страница 142: ...RPM4 AD OPERATION AND MAINTENANCE MANUAL 2005 2007 DH Instruments a Fluke Company Page 132 N NO OT TE ES S...