RPM4-AD™ OPERATION AND MAINTENANCE MANUAL
© 2005-2007 DH Instruments, a Fluke Company
Page 26
THIRD [RANGE] SELECTION: HI Q-RPT, SHORTCUT TO AIRSPEED AT
VARYING ALTITUDE
Selects Hi Q-RPT in absolute, differential
measurement mode (
<Hd>
) in which Hi
– Lo Q-RPT differential pressure
(airspeed) is displayed on top line and
Lo Q-RPT absolute pressure (altitude)
on the bottom line. Units of measure are
default or last selected airspeed and
altitude units.
Select A160K dif Hd
airspeed & altitude
FOURTH [RANGE] SELECTION: HL Q-RPT, SHORTCUT TO ALTITUDE,
BEST UNCERTAINTY
Selects HL Q-RPT (Hi and Lo Q-RPTs in
parallel) with altitude displayed on top
line and rate of change on the bottom
line. SDS is set so that the
TEST(Ps)
port is active. Unit of measure is default
or last selected altitude unit.
Select A160K abs HL
altitude, parallel
FIFTH [RANGE] SELECTION: HL Q-RPT, SHORTCUT TO AIRSPEED, BEST
UNCERTAINTY
Selects HL Q-RPT (Hi and Lo Q-RPTs in
parallel) with airspeed displayed on top
line and altitude corresponding to local
atmospheric pressure on bottom line.
Altitude cannot be changed. SDS is set
so that the
TEST(Pt)
port is active. Unit
of measure is default or last selected
airspeed and altitude units.
Select A160K gage HL
airspeed, parallel
From the fifth
[RANGE]
display, pressing
[RANGE]
again returns to the first
display.
3.3.1.2
RPM4-AD A350K/A160K (FIXED WING)
Pressing
[RANGE]
always displays the active Q-RPT information. Pressing
[RANGE]
again toggles through the other Q-RPT choices available and an air
data shortcut. There are a total of three
[RANGE]
displays.
The sequence of selections when
[RANGE]
is pressed is:
FIRST [RANGE] SELECTION: LO Q-RPT
Selects Lo Q-RPT with all previous
settings for that Q-RPT.
Active A160K Lo
kPa 60g/160a
SECOND [RANGE] SELECTION: HI Q-RPT
Selects Hi Q-RPT with all previous user
settings for that Q-RPT.
Active A160K Hi
kPa 60g/160a
Содержание RPM4-AD
Страница 10: ...RPM4 AD OPERATION AND MAINTENANCE MANUAL 2005 2007 DH Instruments a Fluke Company Page VIII N NO OT TE ES S...
Страница 22: ...RPM4 AD OPERATION AND MAINTENANCE MANUAL 2005 2007 DH Instruments a Fluke Company Page 12 N NO OT TE ES S...
Страница 90: ...RPM4 AD OPERATION AND MAINTENANCE MANUAL 2005 2007 DH Instruments a Fluke Company Page 80 N NO OT TE ES S...
Страница 138: ...RPM4 AD OPERATION AND MAINTENANCE MANUAL 2005 2007 DH Instruments a Fluke Company Page 128 N NO OT TE ES S...
Страница 142: ...RPM4 AD OPERATION AND MAINTENANCE MANUAL 2005 2007 DH Instruments a Fluke Company Page 132 N NO OT TE ES S...