4. REMOTE OPERATION
Page 99
© 2005-2007 DH Instruments, a Fluke Company
READYCK
n
Purpose
Read or set the
Ready
check flag.
Command
Query
“READYCK
n
1”
“READYCK
n
?”
Classic “READYCK
n
=1”
“READYCK
n
”
Optional Suffix
“n”
The active Q-RPT is assumed if no suffix is given.
‘1’ Specify the Hi Q-RPT if the HL RPT is not active
Specify the HL Q-RPT if the HL Q-RPT is active
‘2’ Specify the Lo Q-RPT
‘3’ Specify the HL Q-RPT if the HL Q-RPT is active
Remarks
The internal ready check flag is cleared whenever the specified or active Q-RPT reaches a
Not Ready
(NR) condition. The “READYCK” query returns the status of the flag. The flag is
set by sending the “READYCK 1” program message while the Q-RPT is in a
Ready
condition.
The “READYCK” program message query can then be used at a later time to determine
whether a
Not Ready
condition has occurred since the ready check flag was set.
Example
(enhanced)
Cmd sent:
Query reply:
“READYCK1 1”
“1” (no reply if GPIB-488)
Example
(enhanced)
Cmd sent:
Query reply:
“READYCK1?”
“1”
Example
(enhanced)
Cmd sent:
Query reply:
“READYCK?”
“1” (if Q-RPT condition has stayed
Ready
)
“0” (if Q-RPT condition has NOT stayed
Ready
)
Example
(classic)
Sent:
Query reply:
“READYCK=1”
“READYCK=1”
Example
(classic)
Sent:
Query reply:
“READYCK”
“READYCK=1” (if Q-RPT condition has stayed
Ready
)
“READYCK=0” (if Q-RPT condition has NOT stayed
Ready
)
Errors
ERR# 6:
Argument is not a ‘0’ or a ‘1’.
See Also
“SR”
3.2.4
REMOTE
Purpose
Lock out the front panel keypads during remote operation.
Command “REMOTE”
Classic “REMOTE”
Remarks
The RPM4-AD goes into remote mode whenever communications take place. The user can
return to local operation by pressing the [ESC] key. The REMOTE program message locks
out the front panel completely. The only way to unlock the front panel after the “REMOTE”
command is using the “LOCAL” program message, the IEEE-488 “GTL” command, or by
cycling the RPM4-AD power.
Example
(enhanced)
Cmd sent:
Reply:
“REMOTE”
“REMOTE” (no reply if IEEE-488)
Example
(enhanced)
Cmd sent:
Reply:
“REMOTE?”
“REMOTE”
Example
(classic)
Sent:
Reply:
“REMOTE”
“REMOTE”
See Also
“LOCAL”
Содержание RPM4-AD
Страница 10: ...RPM4 AD OPERATION AND MAINTENANCE MANUAL 2005 2007 DH Instruments a Fluke Company Page VIII N NO OT TE ES S...
Страница 22: ...RPM4 AD OPERATION AND MAINTENANCE MANUAL 2005 2007 DH Instruments a Fluke Company Page 12 N NO OT TE ES S...
Страница 90: ...RPM4 AD OPERATION AND MAINTENANCE MANUAL 2005 2007 DH Instruments a Fluke Company Page 80 N NO OT TE ES S...
Страница 138: ...RPM4 AD OPERATION AND MAINTENANCE MANUAL 2005 2007 DH Instruments a Fluke Company Page 128 N NO OT TE ES S...
Страница 142: ...RPM4 AD OPERATION AND MAINTENANCE MANUAL 2005 2007 DH Instruments a Fluke Company Page 132 N NO OT TE ES S...